Ultisil Hilic NH2

Ultisil Hilic NH 2 is a hydrophilic interaction liquid chromatography (HILIC) column with a specially treated silica matrix to enhance hydrophilicity. It offers improved stability and reproducibility for mixed monosaccharides compared to XB-NH 2 , thanks to the combined effects of weak anion exchange interactions and hydrophilic partitioning forces.

Request a quoteDownload the datasheet

Ultisil Hilic NH2

Description

Ultisil Series HPLC Columns are based on ultra-pure (purity > 99.999%) spherical and totally porous silica, unique bonding chemistry and proprietary surface modification techniques, producing excellent peak shape, column efficiency and exceptional lot-to-lot reproducibility. Ultisil column is the best choice for method development. owing to complete bonding chemistries and stable performance

Features

  • Ultra-pure spherical porous silica, purity > 99.999%
  • Wide pH range: 2.0-8.0
  • High efficiency: theoretical plate > 80000/m
  • Excellent peak symmetry: tailing factor=0.95~1.05
  • Exceptional lot-to-lot reproducibility

Trace Amount Metal Contents Test

A useful chromatographic test of trace amount of metal contents in the column is to compare the peak symmetry of one pair of positional isomers, 4,4’-dipyridyl and 2,2’-dipyridyl, and a neutral chelating reagent, 1,2-dihydroxylnaphthalene. 4,4’-dipyridyl, which cannot form chelating complex with metal, is used as a reference. 2,2’-dipyridyl and 1,2-dihydroxylnaphthalene, which are chelating reagents, are sensitive to trace amount metal in silica. When a C18 column based on type A silica or other so-called type B silica with higher metal content is used, the peaks of 2,2’-dipyridyl and 1,2-dihydroxylnaphthalene would tail or even totally disappear.

ColumnUltisil® XB-C18, 4.6 ×150 mm, 5 μm
Mobile Phase20 mM phosphate(pH 7.6) / methanol=55/45
Flow Rate1.0 mL/min
Detector215 nm
Temperature25°C
Injection Volume1 μL
Samples1) 4,4’-Dipyridyl; 2) 2,2’-Dipyridyl; 3) 1,2-Dihydroxylnaphthalene

Conclusion: Ultisi XB-C18 provides good peak shapes in the separation of these three compounds under pH 7.6, which indicates Ultisil silica contains hardly any metals.

Specifications

Pore size120 Å
Particle size3 µm, 5 µm, 10 µm
ID3.9 mm, 4.6 mm
Length150 mm, 200 mm, 250 mm, 300 mm
PhaseHILIC NH2
USP ClassificationL8
Separation ModeHILIC Mode
pH Range2.0-8.0
Surface Area320m2/g
Carbon Load4%
Max. Pressure60 MPa
EndcappedNo
Max. Temp.60°C(pH≤6.5) and 40°C(pH>6.5)
Particle ShapeSpherical

Configure the instrument

Choose the model, probes and accessories. Incompatible combinations are blocked as you go; the remaining rules are checked when the instrument is added to the quote.