A unit of AKRIBIS One

Xtimate_Bio_SEC-300-Series

Xtimate Bio SEC-300

Xtimate Bio SEC column features ultrapure silica microspheres as matrices, with a uniform surface bonded with a hydrophilic polymer of nanometer thickness. It effectively reduces the adsorption of biological samples on the column, making it ideal for the analysis of proteins, antibodies, cytokines, and other biological samples.

Request a quoteDownload the datasheet

Xtimate Bio SEC-300

Description

Xtimate Bio SEC Chromatography Column employs ultrapure silica gel as its matrix for size exclusion chromatography. Its stationary phase consists of ultrapure silica gel microspheres with a uniformly nano-thick layer of hydrophilic polymer bonded to their surfaces.

Welch Materials employs a unique surface modification technology, ensuring the complete and uniform bonding of the polymer nanolayer to the silica gel surface. This extensive coverage significantly reduces the nonspecific adsorption of biological samples within the silica gel packing. The Xtimate Bio SEC column also exhibits excellent stability and batch reproducibility. Its primary application areas include biomolecules such as proteins, nucleic acids, peptides, oligonucleotides, bacteria, viruses, and more.

Features

  • Stationary phase consists of hydrophilic polymer with a uniform nano-thick layer bonded to the surface of ultrapure silica gel microspheres.
  • Available in 5μm and 3μm silica gel microspheres, catering to high throughput testing requirements.
  • Offers products with different pore size parameters to meet diverse needs for the separation of biomolecules.
  • Minimal nonspecific adsorption of biological molecules.
  • Demonstrates excellent stability and reproducibility across batches.
Xtimate Bio SEC-300

Applications:

1. Bovine serum albumin

Overlap spectra:

Black: Welch Xtimate Bio SEC-120 column

Red: SEC column from a certain manufacturer

Xtimate Bio SEC-300

Conclusion:

Xtimate Bio SEC-120 chromatography column, when used for detecting bovine serum albumin samples, exhibited significantly higher sensitivity and response values compared to a certain manufacturer’s SEC chromatography column. Moreover, it demonstrated lower nonspecific adsorption of samples.

2. Monoclonal Antibody Samples

Xtimate Bio SEC-300

Conclusion:

  • Under these chromatographic conditions, the main peak of the monoclonal antibody sample exhibits excellent peak shape.
  • The separation between the main peak and the preceding unknown peak is satisfactory, meeting the detection requirements.

3. Human Granulocyte Colony-Stimulating Factor Sample

Xtimate Bio SEC-300

Conclusion:

  • Under these chromatographic conditions, the main peak of the human granulocyte colony-stimulating factor sample exhibits excellent peak shape.
  • The separation between the main peak and the surrounding unknown peaks is satisfactory, meeting the detection requirements.

Specifications

Pore size300 Å
Particle size3 µm, 5 µm
ID4.6 mm, 7.8 mm
Length50 mm, 250 mm, 300 mm
PhaseBio SEC-300
USP ClassificationL33 and L59
Separation ModeSize Exclusion
pH Range2.0-8.0
Max. Pressure200Bar
Max. Temp.5-60°C
Particle ShapeSpherical
Column Cross Referenceand may affect the results for certain circumstances. Click here for a more up-to-date cross-reference tool .)

Configure the instrument

Choose the model, probes and accessories. Incompatible combinations are blocked as you go; the remaining rules are checked when the instrument is added to the quote.

Documents

Write on WhatsApp