Frequently Asked Questions
Q: For different samples and gradient conditions, should the Ghost-Buster column be removed or changed?
A: Not necessary. But it needs to be removed only for special circumstances like changing of peak position or ion-pair solvents mobile phase.
Q: When gradient elution changed to isocratic, should the Ghost-Buster column be removed?
A: No need to take the Ghost-Buster column if it did not affect the separation, as the elution of mobile phase stays same under isocratic condition. But impurities in mobile phase shall be taken into consideration.
Q: In gradient system, Ghost-Buster column increases the mixed dwell volume. Will this affect the separation?
A: The packing volume of a 4.6x50mm column is ~400 µL and the column is installed before the injector, which would cause little influence on the analysis. lf it does, connect the Ghost-Buster column to the aqueous phase path before the mixer or switching valve.