Una unidad de AKRIBIS One

LC3600

LC3600 Ultra High Performance Liquid Chromatograph

UHPLC has been gradually adopted in industrial labs, especially the pharmaceutical industry dueto its high resolution, high speed, and solvent saving, since its introduction in early 2004. A UHPLC method using a sub-2 um column could reduce the analysis time by up to 80% and savethe mobile phase consumption by at least 80% compared with an HPLC method using aconventional 3.5 um column without sacrificing separation performance. In addition, the muchshorter run time significantly reduces UHPLC me

Solicitar cotización

LC3600 Ultra High Performance Liquid Chromatograph

Description

Extremely Low Cross-contamination UHPLC Ultra-high Performance High Pressure Liquid Chromatography Equipment 150MPa
Description of Ultra High Performance Liquid Chromatography Instrument

Parameters of Ultra High Performance Liquid Chromatography Instrument

Performance of Ultra High Performance Liquid Chromatography Instrument
1. The dynamic compression compensation of the ultra-efficient binary high-pressure pump adpots high-precision real-time pressure following technology to achieve the effect of "zero fluctuation" and bring extremely high flow stability.

2. The ultra-efficient autosampler adpots ultra-high pressure needle in the flow path non-destructive sampling technology, with a very low sampling cycle period.

3. Needle-in-flow autosampler with ultra-low injection cycle time

4. Extremely low cross-contamination

5. Active preheat function, fast warming rate

LC3600 Ultra High Performance Liquid Chromatograph

Specifications

Work Power220V, 50HZ
Ambient Temperature10-~40°C
Relative Humidity20-85%

System Indicators

Qualitative RepeatiblityRSD ≤0.1%
Quantitative RepeatabilityRSD ≤0.3%

High Pressure Constant Current Pump

Pump Drive ModeLinear Motor Pump
Gradient ModeBinary (4-way solvent)
Flow Setting Range0.0001~2ml/min
Flow Setting Increment0.001ml/min
Max Operating Pressure150MPa
Flow Setting Error±1.1%
Flow Stability≤0.09% RSD
Gradient Stability≤0.02% RSD

Column Oven

Temperature Setting Range20~90°C
Temperature Stability± 0.3°C/h
Error of Temperature Setting Value± 0.5°C

Autosampler

Maximum Withstand Pressure150MPa
Injection Volumn Range0.1~10ul
Injection Volume Error± 0.2ul (Injection Volume 10ul)
Injection Repeatability< 0.3% RSD
LinearityR≥0.999
Cross-contamination< 0.001%
Number of Smples96 Vials (2*48, 2ml)
Temperature Seting Range4~40 °C
Temperature Stability± 1°C
Sampling Speed15s

Ultra-efficient DAD Detector

Baseline Noise≤10 -6 AU
Baseline Drift≤0.9x10-3 AU/h
Minimum Test Concentration2×10-9g/ml (naphthalene/methanol)
Wavelength Range190~800nm
Wavelength Indication Error≤ ±1nm
Wavelength Repeatability≤ ± 0.1nm
Linear2.0 AU
Sampling RateUp to 100Hz
Number of Arrays1024
Light SourceDeuterium (D2) Lamp