GC6100
GC6100 Gas Chromatograph
GC6100 Gas Chromatograph
GC6100 Gas Chromatography with FID Detector
Technical Parameters of GC6100 Gas Chromatography
1.1 Very low noise, FID noise is better than the international leading level;
1.2 High sensitivity detector, low baseline drift;
1.3 Good reproducibility of peak area and retention time;
1.4 High temperature control accuracy and good stability, the column thermostat provides a constant temperature environment for the effective separation and detection of various components.
2.1 Optional multiple detectors (FID, FPD, ECD);
2.2 Any two detectors and two chromatographic columns can be installed at the same time; It can have three independent circuits, gas system and three signal output terminals at the same time;
2.3 Flexible syringe injection system, optional autosampler or six-port valve injection.
Environmental Protection: Trace Toxicants in Air and Water Contaminated Areas Analysis, monitoring and research
Biochemistry: clinical applications, pathology and toxicology studies.

| Power Supply | 220V ±10%, 50Hz ±2% |
|---|---|
| Ambient Temperature | 5°C to 40°C |
| Relative Humidity | 20% to 85% |
| Retention Time repeatability | <0.008% or <0.0008min |
|---|---|
| Peak Area Repeatability | ≤0.5% RSD |
| Injection Port | Split/Splitless |
|---|---|
| Maximum Operating Temperature | 450°C |
| Control Precision | 0.001 psi |
| Control Range | 0–100 psi |
| Gas Control Modes | Constant pressure, constant flow, programmed pressure/flow ramps |
| Maximum Split Ratio | 12,500:1 |
| Optional Carrier Gases | N₂, H₂, He, Ar |
| Flow Rate Range (N₂ as carrier gas) | 0–500mL/min |
| Temperature Range | Ambient +4°C to 450°C |
|---|---|
| Programmable Temperature Ramps | 32 steps/33 platforms |
| Maximum Heating Rate | ≥75°C/min |
| Cooling Time (22°C ambient) | From 450°C to 50°C in <5 minutes |
| Maximum Program Run Time | 9999.99 minutes |
| Maximum Operating Temperature | 450°C |
|---|---|
| Minimum Detectability | ≤1.5pg/s |
| Linear Range | ≥10⁷ |
| DataAcquisition Rate | Up to 1000 Hz (provide proof of software interface) |
| Minimum Detectability | ≤4×10⁻¹⁵ g/mL |
|---|---|
| Linear Rang | ≥10⁴ |
| Sensitivity | ≥60,000 mV·mL/mg |
|---|---|
| Linear Range | ≥10⁴ |
| Instant Noise | ≤0.1 mV |
| Baseline Drift | ≤0.2 mV |
| Sample Capacity | 16 positions |
|---|---|
| Solvent Bottle Positions | 2 |
| Waste Bottle Position | 1 |
| Syringe Cleaning Modes | Standalone, pre-injection, post-injection |
| Injection Volume | 0.1–5μL |